Table 2-4 explains the icons that appear on various screens and their meaning.
Icon
Indicates that the system is controlled by the I/O.
Indicates a feature is locked.
Indicates a turbo pump fault. Turbo pump (Manual Valve screen). If flashing
red on Home screen this indicates a turbo failure.
Indicates that a valve is open. A red filled circle indicates the valve is failing.
Indicates there is a valve fault.
Indicates that a valve is closed.
Selects a graphical representation (xy graph) of leak rate versus time.
ConvecTorr gauge
Numerical representation of the ConvecTorr gauge pressure.
Bar graph representation of the ConvecTorr gauge pressure.
2-10
VS Series Component Leak Detector
Table 2-4 On-Screen Icons
Indicates the status of the internal calibrated leak:
A FLASHING calibrated leak symbol appears.
Lets you know that the calibrated leak is going to expire, this message
begins to appear 10 days prior to the expiration date.
A steady calibrated leak symbol appears, BLUE in color.
Lets you know that the calibrated leak has expired per the date put into
the system at the Internal Calibrated Leak set-up screen.
A steady calibrated leak symbol appears, RED in color.
Lets you know that the calibrated leak temperature sensor has failed or is
disconnected