Internal Calibration Via The Profinet Option; Full External Calibration Via The Profinet Option - Agilent Technologies VS Series Operation Manual

Component leak detector
Hide thumbs Also See for VS Series:
Table of Contents

Advertisement

5.2.5 Internal Calibration via the Profinet Option

Action: Isolate the leak detector from the test system.

5.2.6 Full External Calibration via the Profinet Option

Manual full external calibration (Tune, Offset, & Ext. Calibration)
Action: Place the leak detection test system in the Test/Measurement mode external
calibrated leak on.
Cyclic Input Data:
Cyclic Output Data: Word 0, Parallel_Enable - Bit 6 - 1
Cyclic Output Data: Word 0, Tune In - Bit 3 - pulse 0,1,0
Cyclic Input Data: Word 0, Busy Bit 6 - 1
Cyclic Input Data: Word 0, Tune OK - Bit 0 - 1
Action: Isolate the external calibrated leak from the system.
Cyclic Output Data: Word 1, Select External or Internal Calibration - Bit 10 - (1 = Ext., 0 =
Int.)
Cyclic Output Data: Word 0, Offset - Bit 2 - pulse 0,1,0
Cyclic Input Data: Word 0, Busy Bit 6 - 1
Cyclic Output Data: Word 0, External Cal - Bit 4 - pulse 0,1,0
Cyclic Input Data: Word 0, Busy Bit 6 - 1
Cyclic Input Data: Word 0, Ext Cal OK - Bit 1 - 1
5-8
VS Series Component Leak Detector
Active = 1
NOTE
Cyclic Input Data:
Cyclic Output Data: Word 1, Parallel_Enable - Bit 6 - 1
Cyclic Input Data (External Cal Selected): Word 1, Bit 10 - 0
Cyclic Output Data (Select External or Internal Calibration):
Word 1, Bit 10 - 0
Cyclic Output Data: Word 0 - Internal Calibration - Bit 0 - pulse
Cyclic Input Data:
Word 0, Ready - Bit 5 - 1
Quick external calibration - If a successful internal calibration
NOTE
has been performed it is unnecessary to do an external tuning
procedure since the tuning was done during the internal
calibration. Perform only an Offset and an External Calibration.
Word 0, Ready -Bit 5 - 1
Word 0 - Busy - Bit 6 - 1
Word 0 - Tune_ok - Bit 0 - 1
Word 1 - Interna_Cal ok - Bit 1 - 1