Advertisement

VS Series Component Leak Detector
5. Remove the five (5) Philipps head screws and washers (Figure B-33).
6. Pull the board gently straight up to avoid damaging the pins underneath.
7. Lower the new board into position by carefully aligning the pins.
8. Reinstall the following connections (Figure B-32 on page B-22):
P2 at J1
P4 at J4
9. Reinstall the five (5) Philipps head screws and washers.
10. Use an M4 Allen wrench to reinstall the two screws on the unit's top and open the unit
(Figure B-14 on page B-11).
11. Reconnect the unit and powerup.
12. Ensure that the System Ready Condition filament fault conditions off.
Figure B-33 Mounting Screws
B-23