Transfer Points With Stabilization Wait Timer - Agilent Technologies VS Series Operation Manual

Helium mass spectrometer leak detectors
Hide thumbs Also See for VS Series:
Table of Contents

Advertisement

VS Series Helium Mass Spectrometer Leak Detectors

2.2.4.5.1 Transfer Points with Stabilization Wait Timer

Figure 2-14 shows the Transfer Points screen.
Figure 2-14 Transfer Points with Stabilization Wait Timer
Stabilization Wait
This feature is available November 2010.
NOTE
This appears on the Home screen after a successful startup. This
timer is necessary for the spectrometer stabilization and degassing.
The factory default is
customers who infrequently use the leak detector or expose the
leak detector to aggressive applications to select the 9 or 15 minute
selection. For example: Select 9 minutes for a leak detector that is
used once a week or select fifteen minutes for a leak detector that is
used once a month or less.
5
minutes. It is highly recommended for those
2-29