Appendix B. Replacement Procedures

B.1 VS-C15 Field Replacements: p/n EXVSC15BU
Field exchange replacements of component leak detectors is often more practical where
testing uptime is critical and where troubleshooting time to perform failure and root cause
analysis is a constraint.
Once the leak detector has been programmed (setup) and is ready for testing or is needs to
be taken offline, record the operational parameters. This can be done manually or through
the RS232 port with the DUMP-PARAMS command.
See Section 4 "RS-232 Interface and Operation" on page 4-1.
To perform a dump:
1. Connect to the serial port of the VS-C15.
2. Open HyperTerminal and open the Transfer Menu then select Capture Text and create a
file name in a convenient location.
3. Enter DUMP-PARAMS <CR>
ENABLE-0ENABLE
15 INIT-AUDVOL DISABLE-VOLUME
3.0E-06 INIT-EXTLEAK
1 INIT-INTERNAL-TYPE
1.4E-07 INIT-STDLEAK
25.0 INIT-LEAKTEMP
+0.0 INIT-TEMPFACTOR
1 1 2005 INIT-LKEXPIRE
INTERNAL
LR1REJECT 1.0E-08 INIT-1REJECT DISABLE-1REJECT HIGH-1REJECT
LR2REJECT 1.0E-08 INIT-2REJECT DISABLE-2REJECT HIGH-2REJECT
LR3REJECT 1.0E-08 INIT-3REJECT DISABLE-3REJECT HIGH-3REJECT
LR4REJECT 1.0E-08 INIT-4REJECT DISABLE-4REJECT HIGH-4REJECT
1.0E-08 INIT-AREJECT DISABLE-AREJECT HIGH-AREJECT
-06 INIT-RANGE
-08 INIT-RANGESTOP DISABLE-RANGESTOP
VS Series Component Leak Detector
B-1