7. Pull the board gently straight up to avoid damaging the pins underneath (Figure B-30).
8. Lower the new board into position by carefully aligning the pins seen in Figure B-30.
9. Reinstall the black connector (2140-301) to the Mother board at J306 (
Figure B-29 on page B-20).
10. Reinstall the four (4) Philipps head screws and washers (Figure B-28 on page B-20).
11. Reconnect R2142-301 (white/black) connector to the J100 connector (Figure B-27 on
page B-19).
12. Use an M4 Allen wrench to reinstall the two screws on the unit's top and open the unit
(Figure B-14 on page B-11).
13. Reconnect the unit and powerup.
B.7 Ion Source Board
Equipment Required:
M4 Allen Wrench
Small Philipps Screwdriver
1. Disconnect the power to back of the unit and unplug.
VS Series Component Leak Detector
Figure B-30 Pins Underneath
Disconnect power from the unit before performing any
WARNING
maintenance procedure that requires physically disconnecting
any part of the system.
Static sensitive device, ensure that personnel are properly
CAUTION
grounded before proceeding.
B-21