Gaging onnectors to be Mated with the 85025A/B/D/E
It is important to gage connectors which will be used with the 85025A/B/D/E.
he specications for attaching connectors may vary, depending on the
connector used and the application. However, if the following guidelines are
used, you can avoid connector damage due to tolerance problems.
Type-N female
Gage any female device to be mated with the 85025A. When mated, a type-N
connector pair must have separation between the tip of the female contact
ngers and the shoulder of the male contact pin. Do not use female type-N
connectors that have an inner rubber washer, they damage the male center
conductor.
Precision 7 mm
Gage with the center conductor collet removed, refer to the tolerances given in
the device's manual. here should be no protrusion of the center conductor in
front of the outer conductor mating plane. Check the collet for proper spring
action. Collets should only be removed with a collet extraction tool. Refer to
Microwave Connector Care Quick Reference Card
the
for information.
Precision 3.5 mm female
Gage any female connector to be used with the 85025B/E. here should be
no protrusion of the tip of the female contact ngers in front of the outer
conductor mating plane.
Ag lent 85025A/B/D/E
Detector Ma ntenance
8-3