LCD Display

Power and Static Sensitivity

The power supply requirements for VS C15 Component leak detector: 24 VDC, 5 A min.
The VS Component leak detector can be ordered with or without an external 24 VDC
AC-DC power supply.
When ordered with the universal external AC-DC power supply (VSC15PS), you
must provide a mains supply:
When ordered without the external power supply, you must supply a 24 VDC
+/- 5%, 5 A min. power supply with a local power switch.
Provide the 24 VDC power to the VS Component leak detector via a detachable cable: 18
AWG triple-twisted wires terminated by Molex's HCS-125 Receptacle (P/N 03-12-1036),
with 3-socket female terminals (Molex P/N 18-12-1222) that are compatible with a DC inlet
connector of an EMI filter on the rear panel of the VS C15 Base Unit. Wire as follows:
+24 VDC: position 1 (D-shape terminal) of the receptacle.
PE wire: position 2 (middle terminal) of the receptacle; to provide grounding
between the VS C15 units and the external power supply system.
24VRTN wire: position 3 of the receptacle.
xix
VS Series Component Leak Detectors
If the LCD display breaks, avoid getting any liquid crystal in mouth
WARNING
or eyes. If it gets on the operator hands, feet or clothing, wash it
immediately with soap and water.
Do not apply excessive pressure to the LCD display as this may
CAUTION
cause smears.
In general, the lower the temperature, the longer it takes the LCD
display to turn on. Performance of the LCD display may deteriorate
in low temperature.
100-230 VAC ± 10%, 50/60 Hz ± 2%, with a 10 A local power switch/circuit
breaker.

Advertisement