Reject Set Points
2-26
VS Series Component Leak Detector
Figure 2-8 Reject and Audio Setpoints
There are four possible reject set points in Figure 2-8 on
page 2-26:
Each set point can be programmed to operate around leak
rate or test port pressure (if an optional ConvecTorr gauge is
installed) by pressing the button, which will toggle between
LEAK RATE and PRESSURE.
Change the values of the four Reject Set Points by following
the Data Entry Pad Operation above. Data is in units
specified by the Units screen.
Set points may be turned ON or OFF (enabled or disabled)
for each set point
Each reject set point is set to activate above or below the
programmed value. To activate the reject set point
Above the programmed value, select >.
Below the programmed value, select <.
Active reject set points cause Reject to appear on the HOME screen
along with the number of the reject set point that has been
activated, which blinks to highlight the reject. The status of the
reject set points is also available through the RS-232 line (see
Section 4 "RS-232 Interface and Operation" and Section 5
"Profinet Interface and Operation" on page 5-1).