Appendix C. Troubleshooting

C.1 Service 1/Service 2 Commands
The Service1 and Service2 RS232 commands provide a detailed setup and operational
status of the VS-C15 component leak detector, especially for customers who do not have
optional Front Panel Display. This information is very helpful in diagnosing problems in the
field for Agilent's Service and Technical Support Representatives.
C.1.1 SERVICE1
Entering this command elicits a response from the VS-C15 as follows:
VS Series Component Leak Detector
Current Date: 1 11/03/2008 12:10:26.95 1
IONCHAMBER 244.6
EMISSION 0.553
OFFSET (Sys/Std/Curr) -2914 0 -2914
GAIN (Sys/Std/Curr) 1.0 1.0 1.0
LPV: 0.01E-09 008426 000001 CONTRAFLOW
Turbo Speed (RPM): 33000
range -06 auto
exponent -06
reject 1.0E-08 1.0E-08 1.0E-08 1.0E-08 1.0E-08
stdleak 1.4E-07
output
linear
filament Two Lit internal
Filament 1 Hours: 345
Filament 2 Hours: 0
Turbo Ready
Turbo No Fault
Turbo Speed (RPM): 33000
Turbo Temp (Celsius): 30
Turbo Life (Hrs): 355
STD_LEAK_VALVE Closed
Current Cal Leak Temperature: 24.5
Internal Calibration OK?: No
Internal Calibration Error: Lost-
Power
External Calibration OK?: No
C-1