Automatic configuration with Bootp

When automatic configuration with Bootp is selected and the LAN interface is
powered on, it broadcasts a BOOTP (Bootstrap Protocol) request that contains
its MAC (hardware) address. A BOOTP server daemon searches its database
for a matching MAC address, and if successful, sends the corresponding
configuration parameters to the detector as a BOOTP reply. These parameters
become the active TCP/IP parameters immediately and the TCP/IP connection
can be established.

Configuring the Agilent Bootp service program

All examples shown in this chapter will not work in your environment. You
N O T E
need your own IP-, Subnet-Mask- and Gateway addresses.
Assure that the detector configuration switch is set properly. The setting
N O T E
should be either Bootp or Bootp & Store, see
Assure that the detector connected to the network is powered off.
N O T E
If the Agilent Bootp Service program is not already installed on your PC, then install it from
N O T E
your Agilent ChemStation CD-ROM, located in folder \Bootp. The screens refer to version
B.01.0x
40
Table 6
on page 35.
1100 Series DAD and MWD User Manual