20. On the PCB, use the resistance meter to verify an open circuit between any two of the
six ion source header pins (except FIL-1 to FIL-1 and FIL-2 to FIL-2 which should be 0.3
Ohms or less). Also verify an open circuit between the body of the spectrometer and any
of the ion source header pins (Figure B-87). If there is continuity (short circuit) at any of
the points, remove the header and inspect for shorting.
21. Re-Install the cover with the four screws.
22. Re-plug in both cables and ground strap.
23. Verify that turbo vent valve is closed.
24. Leak check the spectrometer to ensure a leak free joint.
25. Close the cover and secure using existing hardware.
26. Power up the leak detector.
27. Let the unit run for 20-30 minutes, then perform a calibration (via the I/O, front panel
display or RS232), per the user manual, to validate a successful installation.
B-58
VS Series Component Leak Detector
Figure B-87 Ion Source Header Pin Schematic