2.2.5.1 Calibration Set-Up Screen

Press Calibration Leak Set-Up on the Menus screen.
The Calibration Set-Up screen appears (Figure 2-7). Use this screen to:
Internal Leak Value
VS Series Component Leak Detector
View the value and expiration date of the Internal Calibrated Leak
Enter the External Calibrated leak value
Select whether to use an Internal or external calibrated leak
Figure 2-7 Calibration Set-up Screen
If the internal calibrated leak option is not installed, the internal
NOTE
leak value and expiration date is grayed out and the internal
leak selection is inactive.
An optional calibrated helium leak can be installed inside the leak
detector unit and used to perform a FULL INTERNAL
CALIBRATION of the leak detector. The temperature-compensated
internal leak value and expiration date appear to the right of the
Internal Leak Value text if this option is selected. The actual
Internal Leak Value and Expiration Date can be modified using the
Internal Calibration Leak screen, which is located in the
Maintenance screen.