Discrete I/O Board Wiring; Profinet And I/O Connectors - Agilent Technologies VS Series Operation Manual

Component leak detector
Hide thumbs Also See for VS Series:
Table of Contents

Advertisement

4. Disconnect (Figure B-39):
1. P301 (Red/Black) J301
2. PG1_2
3. P300
5. Remove the (Figure B-40):
Two (2) Philipps screws holding the Profinet connector.
Two nuts holding the I/O connector, using a 5 mm nut driver.
6. Remove the old card and slip the new card in place.
7. Reconnect (Figure B-39 on page B-28):
P301 (Red/Black) J301
PG1_2
P300
8. Reinstall the (Figure B-40):
Two (2) Philipps screws holding the Profinet connector.
Two nuts holding the I/O connector, using a 5 mm nut driver.
9. Use an M4 Allen wrench to reinstall the two screws on the unit's top and open the unit
(Figure B-14 on page B-11).
10. Reconnect the unit and powerup.
B-28
VS Series Component Leak Detector
1
Figure B-39 Discrete I/O Board Wiring
1
2
Figure B-40 Profinet and I/O Connectors
2
3
3