Inside Of Spectrometer Body; Coupler Screws - Agilent Technologies VS Series Operation Manual

Component leak detector
Hide thumbs Also See for VS Series:
Table of Contents

Advertisement

10. If the inside of the spectrometer body appears to be dirty, clean it with a lint free wipe
and Isopropyl alcohol (Figure B-82).
NOTE
11. On a clean work surface, remove the defective filament(s) from the spectrometer by
loosening the two coupler screws. Then loosen the M2 screw and washer and remove
the defective filament (Figure B-83).
Filament Assembly
VS Series Component Leak Detector
Agilent recommends that you use the Vac-u-solv spectrometer
cleaning kit (PN: 670029096).
Figure B-82 Inside of Spectrometer Body
Figure B-83 Coupler Screws
Filament Alignment
Surfaces
Coupler
B-55