Home Screen - Agilent Technologies VS Series Operation Manual

Component leak detector
Hide thumbs Also See for VS Series:
Table of Contents

Advertisement

2.2.3 HOME Screen

The HOME Screen is displayed on Power-Up with delay about 10 seconds.
The HOME Screen displays the leak rate in the linear and log bar graph and digital format,
a summary of the leak detector status and configuration settings:
Linear Bar Graph
Digital Leak Rate Display
Leak Detector Status (Table 2-
5 on page 2-13)
Reject Status Indicator
Control Panel button
Two soft key buttons, Control Panel and Menus, give access for changing the leak detector
configuration (Figure 2-2).
Linear Bar Graph
Display
Log Bar Graph Display
VS Series Component Leak Detector
Log Bar Graph
Test Port Pressure
Condition Indication (Table 2-
6 on page 2-14)
Menus button
Figure 2-2 HOME Screen
Displays one decade of the leak rate reading in linear mode and
range exponent in digital format
Displays full range of the leak rate reading in logarithmic mode
Zero button
2-11