Mother Board Wiring; Mother Board Mounting Screws - Agilent Technologies VS Series Operation Manual

Component leak detector
Hide thumbs Also See for VS Series:
Table of Contents

Advertisement

VS Series Component Leak Detector
7
6
8
5
1
2
9
10
4
3
Figure B-36 Mother Board Wiring
8. Remove the six (6) Philipps screws holding the board (Figure B-37).
Figure B-37 Mother Board Mounting Screws
9. Remove the board and mount the new board with the six (6) Philipps screws.
10. Reconnect the wiring as in (Figure B-36 on page B-26).
11. Reinstall the nine (9) sets of spacer hardware that separated the smaller boards from the
Mother board (Figure B-35 on page B-25).
12. Reinstall the Ion Source board as per Section B.7 "Ion Source Board" on page B-21.
13. Reinstall the board as per Section B.6 "CPU (Brain) Board" on page B-18.
B-26