14. Validate that the correct internal leak data was input (data from the Calibration
Certificate):
15. Perform an internal calibration (via the I/O, Front Panel Display or RS232) to validate a
successful installation of the internal calibrated leak.
16. Leak check the calibrated leak module to ensure a leak free joint between the
atmosphere and vacuum space inside.
17. Attach the cover and secure using existing hardware.
B.5 Turbo Controller Board
Equipment Required:
M4 Allen Wrench
Small Philipps Screwdriver
1. Disconnect the power to back of the unit and unplug.
2. Wait 30 seconds for the high voltage to dissipate.
B-16
VS Series Component Leak Detector
?STDLEAK (reports back the value of the internal calibrated leak).
?LEAKTEMP (reports back the temperature of the calibrated leak from the factory
calibration).
?TEMPFACTOR (reports back the temperature coefficient).
?LKEXPIRE (reports back the date of expiration).
Disconnect power from the unit before performing any
WARNING
maintenance procedure that requires physically disconnecting
any part of the system.
Static sensitive device, ensure that personnel are properly
CAUTION
grounded before proceeding.