VS Series Component Leak Detector
7. Using an M5 allen wrench, remove the six screws holding on the spectrometer header
assembly (Figure B-62).
Figure B-62 Spectrometer Header Assembly Screws
NOTE
8. Remove the spectrometer header assembly (Figure B-63).
Figure B-63 Spectrometer Assembly Removal
Wear non-powdered, ESD-safe Nitride or equivalent gloves (not
included in kit) to prevent skin oils from getting on vacuum
surfaces. Ensure that personnel are properly grounded before
proceeding.
B-43