Lamp Current too Low
Probable cause
1
2
3
Wait Function Timed Out
Wait for temperature or wait for defined signal has not been fulfilled within
the specified time frame.
Probable cause
1
Agilent 1260 Infinity RID User Manual
Optical unit cable disconnected.
Defective main board.
Defective lamp or optics.
Time too short.
Refractive Index Detector Specific Error Messages
Suggested actions
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Suggested actions
Increase time.