VS Series Component Leak Detector
Figure B-22 Motherboard Connection
10. Power up the VS C15 component leak detector.
11. Set up the Internal Calibrated Leak Option via the front panel display by:
a. Select Advanced Parameters > Internal Type and select STD LEAK.
b. Select Maintenance > Internal Calibrated Leak and input the calibration data from
the Calibration Certificate for the internal calibrated leak including:
Internal Leak Value
Temperature
Temperature coefficient factor
Cal Leak - Date of Expiration
c. Validate the setup using the Calibration Set Up menu: the temperature compensated
leak value, internal calibrated leak temperature and date of expiration.
12. Set up the Internal Calibrated Leak Option via the RS232 serial communications port
(see the operation manual for communicating via the RS232 serial port).
13. Enter the following commands:
1 INIT-INTERANL-TYPE
X.XE-0X INIT-STDLEAK (Enter the leak rate value from the calibration certificate
ex. 1.8E-07 INIT-STDLEAK).
XX.X INIT-LEAKTEMP (Enter the temperature from the calibration certificate ex.
23.5 INIT-LEAKTEMP).
-/+XX INIT-TEMPFACTOR (Enter the temperature coefficient from the calibration
certificate ex. -7 INIT-TEMPFACTOR).
mm dd yyyy INIT-LKEXPIRE (Enter the date of expiration from the calibration
certificate ex. 12 22 2008 INIT-LKEXPIRE).
B-15