Configuring A Field Replacement Or New Vs-C15; Turbo Pump Replacement - Agilent Technologies VS Series Operation Manual

Component leak detector
Hide thumbs Also See for VS Series:
Table of Contents

Advertisement

-06 INIT-EXPONENT AUTO
0 INIT-AZ<0
1 INIT-TC-GAUGE
INIT-LINEAR
0 INIT-DISPLAY
777777 6 INIT-PASSWORD DISABLE-PASSWORD
0 INIT-LANGUAGE
4. Open the Transfer Menu and select Capture Text and Stop.
5. Open the created file and delete the first and last line of text file and Save.
B.1.1 Configuring a Field Replacement or New VS-C15
To do this:
1. Open HyperTerminal
a. Select Properties > Settings > ASCII Setup > ASCII Sending: Character delay.
b. Set Input to 200 milliseconds and select Save settings.
2. From the HyperTerminal – Transfer Menu:
a. Select Send Text File.
b. Select the original DUMP-PARAMS file for that leak detector.
3. Verify the saved Dump Parameters Settings against the settings on the front panel display
in the various menus (Control Panel, Calibration Set Up, Reject & Audio Set point Setup,
Leak Rate Ranging, Advanced Parameters, Output Control & Maintenance).
B.2 Turbo Pump Replacement
Equipment Required:
M4 Allen Wrench
Powder-free Nitrile Gloves
10 mm Nut Driver
Adjustable Wrench
B-2
VS Series Component Leak Detector
Disconnect power from the unit before performing any
WARNING
maintenance procedure that requires physically disconnecting
any part of the system.