Section 5. Profinet Interface and Operation

5.1 Protocol (PROFINET-I/O)

5.1.1 PROFINET IO Interface Introduction

The Network I/O PCB implements a PROFINET IO interface using the
®
Anybus
-CompactCom PROFINET IO module. The module exchanges cyclic data on the
ProfiNet network as PROFINET input data (data from the module to the ProfiNet host - PLC
or PC), and PROFINET output data (data from the ProfiNet host to the module). This
document describes the structure and meaning of the exchanged data.

5.1.2 PROFINET I/O Interface Setup

The PROFINET I/O interface is installed into the Controller (PLC or PC) using the GSDML
file located on the CD with this document. The file has the naming convention
GSD-VX.X-Agilent-PISAIO-PRT-YYYYMMDD.XML, where X.X is the revision and
YYYYMMDD is the release date. Also on the CD is a bitmap logo file that is sometimes used
by the development software to identify the VS C15 leak detector. Typically, the GSDML file
(and the bitmap file) are loaded into the development software that is used to write code in
the Controller. Additionally, the programmer might be required to configure further interface
details. If so, the following information may be useful: The interface data is arranged as 8
16-bit Output words followed by 8 16-bit Input words.
VS Series Component Leak Detector
The GSDML File is located on the CD that shipped with the
NOTE
VS-C15.