2.3.14.3 Cal Leak Expiration Set-Up

The Cal Leak Expiration button allows you to put in the calibrated leak expiration date
located on the cal leak label.
1. Press the button next to the Cal Leak Expiration text on the Internal Calibrated Leak
screen.
The Cal Leak Expiration Set-Up screen appears (Figure 2-26).
The Cal Leak Expiration is usually one year from leak installation.
As the calibrated leak approaches its expiration date (starting within 30 days of
expiration), a cal leak symbol will flash on the HOME screen. When the calibrated leak
expires, a symbol appears on the HOME screen steady and blue.
To use this screen:
1. Press the button next to Year or Day and a keypad screen appears to set the value.
2. Press Done for each item configured.
3. Press the button next to Month, use the up and down arrows to change the value and
press OK.
4. Press DONE to set the calibrated leak expiration date and exit the screen.
2-54
VS Series Component Leak Detector
Figure 2-26 Cal Leak Expiration Set-Up