Vs C15 Startup Scenario - Agilent Technologies VS Series Operation Manual

Component leak detector
Hide thumbs Also See for VS Series:
Table of Contents

Advertisement

Table 5-1 Configuration and Status Variables (Continued)
Index
Variable
Access
34
OFFSET
R
35
IONCHAMBER R
36
EMISSION
R
37
FIL1HOURS
R
38
FIL2HOURS
R
39
TURBOHRS
R
System Startup Function
Power up unit: 24 VDC
Turbo ramps to programmed RPM
Turbo reaches programmed RPM
Filaments 1 & 2 are checked and selected
filament activates
System background and turbo bearing
temperature are measured. If these
parameters meet the startup limits the 15
minute stabilization timer stops or times
out.
VS Series Component Leak Detector
Description
Reports the offset number.
Reports the Ion Chamber Voltage.
Reports the Emission Current.
Reports the running hours of Filament 1.
Reports the running hours of Filament 2.
Reports the running hours of th Turbo Pump.
Table 5-2 describes the startup scenario for the ProfiNet
NOTE
interface option.
Table 5-2 VS C15 Startup Scenario
If the system was recently calibrated before shut down, it can
NOTE
be used for qualitative testing. Read the standard leak to verify
accuracy. For more accurate quantitative leak values, perform a
calibration 30 minutes after start-up.
To accurately read in the most sensitive decades of each
working range, perform a tune, calibration and offset, and
internal or external calibration after a 20-30 minute warm up.
The VS C15 Base unit is always in Test/Measurement mode.
NOTE
Word 5: Hi Byte, Word 6: Lo Byte, Word 7: 1 = neg
Word 5: xxxx (Volts x 10)
Word 5: (mA x 1000)
Word 5: Hi Byte, Word 6: Lo Byte, Word 7: 1 = neg
Word 5: Hi Byte, Word 6: Lo Byte, Word 7: 1 = neg
Word 5: Hi Byte, Word 6: Lo Byte, Word 7: 1 = neg
ProfiNet
Cyclic Input Data, Word 0, Ready – Bit 5 –
0
Busy - Bit 6 – 1
Cyclic Input Data, Word 0, Ready – Bit 5 –
0
Busy - Bit 6 – 1
Cyclic Input Data, Word 0, Ready – Bit 5 –
1
Busy - Bit 6 – 0
Response
5-7