B.11 Spectrometer Header M1 Replacement
Equipment Required:
Metric Allen Wrench (Range 1.5mm to
10mm)
Non-powdered ESD-safe Nitrile or
equivalent gloves
Torque wrench, M5 head (must be
adjustable to 45 in-lbs (5.0 N-m) and 90 in-
lbs (10.2 N-m)
1. Disconnect the power to back of the unit and unplug.
2. Wait 30 seconds for the high voltage to dissipate.
3. Use an M4 Allen wrench to remove the two screws on the unit's top and open the unit
(Figure B-46).
VS Series Component Leak Detector
Disconnect power from the unit before performing any
WARNING
maintenance procedure that requires physically disconnecting
any part of the system.
Inspect the new header kit for deformation, damage and
NOTE
filament coating flaking prior to installation. Do not install if
damaged. Touching the filament under any condition will
cause damage.
Figure B-46 Unit Top Screws
M3 Philips head screwdriver
Isopropyl alcohol
B-33