VS Series Component Leak Detector
6. Remove the assembly and use a flat head screw driver to remove the two screws
holding the calibrated leak (Figure B-44).
Figure B-44 Calibrated Leak Screws
7. Install the new calibrated leak using the same two screws.
8. Reattach the leak assembly to the flange with the clamp.
9. Connect the calibrated leak temperature cable end labeled P1A to the leak assembly
board connector J1A (Figure B-45).
Figure B-45 Leak Assembly Board Connection
10. Close the cover and secure using the existing hardware.
B-31