VS Series Component Leak Detector
6. Using an M3 Philips head screw driver, remove the four screws holding the
spectrometer cover and ground strap on (Figure B-78).
Figure B-78 Spectrometer Cover Screws
7. Using an M5 allen wrench, remove the six screws holding on the spectrometer header
assembly (Figure B-79).
Figure B-79 Spectrometer Header Assembly Screws
NOTE
Wear non-powdered, ESD-safe Nitride or equivalent gloves (not
included in kit) to prevent skin oils from getting on vacuum
surfaces. Ensure that personnel are properly grounded before
proceeding.