Calibrated Leak - Agilent Technologies VS Series Operation Manual

Component leak detector
Hide thumbs Also See for VS Series:
Table of Contents

Advertisement

2.1.2 Calibrated Leak

2.1.2.1 Internal calibration via the Front Panel Display
Perform an internal calibration (if installed) to verify that the leak detector is functional and
is ready for testing
Action: Isolate the leak detector from the test system.
1. From HOME screen press MENUS > CALIBRATION SET-UP calibrate with > Internal
Leak.
2. Press Control Panel on the HOME screen and the Control Panel appears.
3. Press Calibrate and the following messages appears on the display:
Calibration Prep
Offsetting
Std Leak Prep
Calibrating
Zeroing
Test
4. Press CONTROL PANEL > STD LEAK > HOME and verify that the VS C15 Base unit
reads the Calibrated Leak correctly. The VS C15 Base Unit performs an automated FULL
internal calibration to the optional installed Calibrated Leak.
2-4
VS Series Component Leak Detector
Verify that the correct information for the internal calibrated
NOTE
leak has been entered in the Maintenance screen, which is
found in the Set Up Menu.