Verify voltage measurement accuracy for each of the voltages listed in Table 18-2. For
each test point:
Select the correct Model 6514 measurement range.
Set the calibrator voltage to the indicated value.
Verify that Model 6514 voltage reading is within the limits given in the table.
9.
Repeat the procedure for negative source voltages with the same magnitudes as those
listed in Table 18-2.
Table 18-2
Voltage measurement accuracy reading limits
Model 6514 range
Calibrator voltage
2V
2.00000V
20V
20.0000V
200V
200.000V
Performance Verification
Model 6514 volts reading limits
(1 Year, 18˚C-28˚C)
1.99946 to 2.00054V
19.9947 to 20.0053V
199.877 to 200.123V
18-11