During zeroing, the ADR-1500 also performs diagnostics on the optics and
flow rates. The only allowable fault is "Background High". All other faults
will result in a failure to zero the instrument. All faults are reported with a
hex coded error flags. The faults can be determined from the following
Table
6–2, but if a fault persists, the ADR-1500 instrument requires
service.
Table 6–2. Troubleshooting – Error Code Responses from Zeroing
Hex code [Flags]
Binary
01 x
0000 0001 b
02 x
0000 0010 b
04 x
0000 0100 b
08 x
0000 1000 b
10 x
0001 0000 b
20 x
0010 0000 b
40 x
0100 0000 b
80 x
1000 0000 b
Fault
Background High
Background calculation FAILED
Source current low
Source current high
Ref. detector low
Ref. detector high
Pump voltage low
Pump voltage high
MIE ADR-1500 Instruction Manual
Troubleshooting
6-5