Figure B-1. Flags; Alarms - Thermo Scientific 42i Instruction Manual

Chemiluminescence no-no2-nox analyzer
Hide thumbs Also See for 42i:

Advertisement

Thermo Fisher Scientific
Figure B–1. Flags
alarm chamber temp min

Alarms

alarm chamber temp max
These commands report the chamber temperature alarm minimum and
maximum value current settings. The example below reports that the
chamber temperature alarm minimum value is 47.0 °C.
Send:
Receive:
set alarm chamber temp min value
set alarm chamber temp max value
These commands set the chamber temperature alarm minimum and
maximum values to value, where value is a floating-point number
representing chamber temperature alarm limits in degrees C. The example
below sets the chamber temperature alarm maximum value to 50.0 °C.
Send:
Receive:
alarm chamber temp min
alarm chamber temp min 47.0 deg C
set alarm chamber temp max 50.0
set alarm chamber temp max 50.0 ok
C-Link Protocol Commands
Model 42i Trace Level Instruction Manual
Alarms
B-13