Maintenance

Flushing the nebulizer - weekly

Flushing the nebulizer - weekly

Passing the vapors of solvents like IPA, methanol, acetonitrile, or water through
the LC/MS system is called steam cleaning and is used to clean possible
contamination in the LC/MS system. The usual conditions are:
Make sure that the MS stream selection valve is set to go to the MS. Steam
Cleaning overnight has been shown to be one of the most important factors in
improving the signal-to-noise level by reducing the contaminations.
To clean an MS, there are two options:
The IPA/MeOH/Water combo can be anything. IPA, maybe 50 (IPA):25:25. IPA
cleans both aqueous and organic.
Schedule
Check the nebulizer weekly. If needed, flush traces of samples and buffers out of
the tubing, valves, and nebulizer.
Equipment List
Steps
1 Install a mix of 30 to 70% methanol in water into a LC solvent bottle.
2 Remove the column from the LC column compartment.
3 Remove nebulizer from spray chamber and support the nebulizer so that it
4 Control the LC flow through the acquisition software to pull from the solvent
5 Set the flow rate to not more than 1 mL per minute.
88
• LC pump flow to 0.5 ml/min
• Nebulizer pressure to 60 psi
• Drying gas to 13 L/min
• Drying gas temp to 350 °C
• Plain water/MeOH at high temp – 50:50 proportion
• IPA/MeOH/Water at high temperature
• LC/MS-grade methanol
• LC/MS-grade water
sprays into an appropriate waste container (beaker, bottle, etc.).
bottles.
Agilent Revident LC/Q-TOF System User Guide