If The Self-Test Fails To Run - Agilent Technologies 8511B Operating And Service Manual

Frequency converter test set
Table of Contents

Advertisement

Table 6-3
Self-Test Indication
PWON
Start ROM Test
Pass ROM Test
Start RAM Test
Pass RAM Test
If the Self-Test Fails to
Run
Self-Test Failure Indications
A4 GPIB LEDs
LSN
TLK
ON
ON
OFF
ON
OFF
OFF
OFF
OFF
OFF
OFF
If the portion of memory which contains the self-test programming is faulty,
the self-test will not run properly. The following conditions indicate that the
GPIB board is most probably faulty.
all LEDs flash briefly and go off
all LEDs flash briefly and stay on
ACTIVE LED goes on too soon
ACTIVE LED does not go on
SRQ
REM
ON
ON
ON
ON
ON
ON
OFF
ON
OFF
OFF
Agilent 8511B Test Set Operating and Service Manual 6-9
Test Set Troubleshooting
Troubleshooting Procedures
Time (after turn-on)
0 to 0.5 seconds
on briefly
0.5 to 2.0 seconds
after 2 seconds