Temperature Compensation Coefficient; Reference Temperature - Thermo Scientific Eutech EC 1720 User Manual

Conductivity meter
Table of Contents

Advertisement

3.2.7 Temperature Compensation
Coefficient
Set temperature compensation off or enter the value of the
linear coefficient value for temperature compensation to
apply a constant percent correction factor to every degree
change in temperature.
To set the temperature compensation off, set the value to
0.0 %/°C.
To set a linear temperature compensation value, enter a
value of 0.0 to 10.0 %/°C.
Press the  or  key to scroll through the Setup
Menu list until the Temperature Compensation
Coefficient item is shown.
Press the  or  key to set the coefficient value.
a. Press the  key to increase the value by 0.01.
Press and hold the  key to quickly increase.
b. Press the  key to decrease the value by 0.01.
Press and hold the  key to quickly decrease.
Press the "Enter" key to save the change.
18
Chapter 3 | Setup Menu

3.2.8 Reference Temperature

Set the reference temperature for temperature
compensation to 15°C, 20°C, 25°C or 30°C.
Press the  or  key to scroll through the Setup
Menu list until the Reference Temperature item is
shown.
Press the  or  key to set the reference
temperature value.
Press the "Enter" key to save the change.