Table 3-4
Auto source delay
2400
2410
I-range
I-range
1µA
1µA
10µA
10µA
100µA
100µA
1mA
1mA
10mA
20mA
100mA
100mA
1A
1A
*10A range only on Model 2430.

Front panel source delay

To set the manual source delay from the front panel:
1.
Press CONFIG then SOURCE V.
2.
Select DELAY from the displayed choices, then press ENTER.
3.
Enter the desired DELAY value, then press ENTER.
4.
Press EXIT to return to normal display.
To set the auto source delay state from the front panel:
1.
Press CONFIG then SOURCE V.
2.
Select AUTO DELAY from the displayed choices, then press ENTER.
3.
Select ENABLE or DISABLE as desired, then press ENTER.
4.
Press EXIT to return to normal display.
Remote command source delay
Use the :SOURce:DELay or :SOURce:DELay:AUTO commands to program the source
delay via remote. (See
command sets the source delay to 500ms:
:SOUR:DEL .5
Similarly, send the following command to enable auto delay:
:SOUR:DEL:AUTO ON
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Basic Source-Measure Operation
2420
2425/2430
2440
I-range
I-range
I-range
10µA
10µA
10µA
100µA
100µA
100µA
1mA
1mA
1mA
10mA
10mA
10mA
100mA
100mA
100mA
1A
1A
1A
3A
3A/10A*
5A
Section
18,
"Set
delay," for details.) For example, the following
3-13
Auto delay
Auto delay
(Source V)
(Source I)
3msec
3msec
2msec
1msec
1msec
1msec
1msec
1msec
1msec
1msec
1msec
1msec
1msec
2msec