Models 2500 and 2502 User's Manual

Set delay

NOTE
DELay <n>
:SOURce[1]:DELay <n>
:SOURce2:DELay <n>
Parameters
Query
Description
AUTO <b>
:SOURce[1]:DELay:AUTO <b>
:SOURce2:DELay:AUTO <b>
Parameters
Query
Description
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
All SOURx:DEL commands are global and affect both simultaneously.
<n> =
0 to 999.998
MINimum
MAXimum
DEFault
:DELay?
:DELay? DEFault
:DELay? MINimum
:DELay? MAXimum
This command is used to manually set a delay (settling time) for the
source. After the programmed source is turned on, this delay occurs to
allow the source level to settle before a measurement is taken. Note that
this delay is the same for both channels.
Do not confuse this source delay with the trigger delay. The source
delay is part of the device action (SDM cycle) while the trigger delay
occurs before the device action. See
panel operation),"
Auto delay can instead be used to automatically set the source delay.
See the next command.
<b> =
0 or OFF
1 or ON
:AUTO?
These commands are used to enable or disable auto delay. When
enabled, the instrument will automatically select a delay period. *RST
and SYST:PRES default is OFF. Note that this delay is the same for
both channels.
SCPI Command Reference
Manually set source delay
Manually set source delay
Specify delay in seconds
0 seconds
999.998 seconds
0.001
Query delay
Query *RST default delay
Query lowest allowable delay
Query highest allowable delay
Section
for more information.
Enable/disable auto source delay
Enable/disable auto source delay
Disable auto delay
Enable auto delay
Query state of auto delay