18-108 SCPI Command Reference

Select guard mode

GUARd <name>
:SYSTem:GUARd <name>
Parameters
<name> =
Query
:GUARd?
Description
This command is used to select the guard mode. OHMS guard is a low-
impedance guard drive used for in-circuit resistance measurements.
Ohms guard is not available for SourceMeters that have 1A, 3A and/or
10A ranges (source or measure). CABLE guard provides a high-
impedance guard drive that is used to eliminate leakage currents in
cabling and test fixtures.
When performing 6-wire ohms guard measurements, use the GUARD
out-put state. The OUTPut [1]:SMODe GUARd command is used to
select the GUARD output-off state.
NOTE
See
Section 2
Initialize memory
INITialize
:SYSTem:MEMory:INITialize
Description
When this command is used, the following actions to initialize battery
backed RAM occur:
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
2400 Series SourceMeter
Select guard mode
OHMS
Ohms guard mode
CABLe
Cable guard mode
Query guard mode
for details on guarding.
Initializes battery backed RAM
TRACe (data store) data is lost, buffer size is reset to 100, and
timestamp is set to the absolute format.
SOURce1:LIST:CURR and VOLT are reset to 0A and 0V,
respectively.
Deletes all user-defined math expressions. (See "CALCulate[1]"
subsystem.)
All 100 memory locations for a memory sweep are initialized to the
present setup configuration of the SourceMeter with CALC 1 dis-
abled. User-defined math expressions are replaced with the
"Power" math expression.
The four standard save setups (*SAV0 - *SAV4) are initialized to
the present setup configuration of the SourceMeter.
All CALCulate1 user-defined math expressions are deleted.
®
User's Manual