Measurements

Average

Average is the sum of the levels of the waveform samples divided by the
number of samples.
Where x
measurement interval.
The Full Screen measurement interval variation measures the value on all
displayed data points.
The N Cycles measurement interval variation measures the value on an
integral number of periods of the displayed signal. If less than three edges
are present, the measurement shows "No edges".
The X cursors show what interval of the waveform is being measured.

DC RMS

DC RMS is the root- mean- square value of the waveform over one or more
full periods.
RMS (dc) =
Where x
measurement interval.
The Full Screen measurement interval variation measures the value on all
displayed data points.
252
Preshoot
local Maximum
local Minimum
Preshoot
=
= value at ith point being measured, n = number of points in
i
2
=1
= value at ith point being measured, n = number of points in
i
Agilent InfiniiVision 4000 X-Series Oscilloscopes User's Guide
Top
Base