Oscilloscope Bandwidth Required - Agilent Technologies 4000 X-Series User Manual

Infiniivision 4000 x-series
Hide thumbs Also See for 4000 X-Series:
Table of Contents

Advertisement

12
Acquisition Control
An oscilloscope's rise time is not the fastest edge speed that the
oscilloscope can accurately measure. It is the fastest edge speed the
oscilloscope can possibly produce.

Oscilloscope Bandwidth Required

The oscilloscope bandwidth required to accurately measure a signal is
primarily determined by the signal's rise time, not the signal's frequency.
You can use these steps to calculate the oscilloscope bandwidth required:
1 Determine the fastest edge speeds.
2 Compute the maximum "practical" frequency component.
3 Use a multiplication factor for the required accuracy to determine the
Required accuracy
20%
10%
3%
See Also
Choosing an Oscilloscope with the Right Bandwidth for your
Application, Agilent Application Note 1588
("http://cp.literature.agilent.com/litweb/pdf/5989-
218
You can usually obtain rise time information from published
specifications for devices used in your designs.
From Dr. Howard W. Johnson's book, High- Speed Digital Design – A
Handbook of Black Magic, all fast edges have an infinite spectrum of
frequency components. However, there is an inflection (or "knee") in the
frequency spectrum of fast edges where frequency components higher
than f
are insignificant in determining the shape of the signal.
knee
f
= 0.5 / signal rise time (based on 10% - 90% thresholds)
knee
f
= 0.4 / signal rise time (based on 20% - 80% thresholds)
knee
oscilloscope bandwidth required.
Oscilloscope bandwidth required
f
BW
f
BW
f
BW
Agilent InfiniiVision 4000 X-Series Oscilloscopes User's Guide
= 1.0 x f
knee
= 1.3 x f
knee
= 1.9 x f
knee
5733EN.pdf")

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents