10
Triggers
Standard
Generic
EDTV 480p/60
EDTV 567p/50
HDTV 720p/50
HDTV 720p/60
HDTV 1080p/24
HDTV 1080p/25
HDTV 1080p/30
HDTV 1080p/50
HDTV 1080p/60
HDTV 1080i/50
HDTV 1080i/60
7 Press the Auto Setup softkey to automatically set up the oscilloscope for
190
The Generic selection lets you trigger on custom bi- level and tri- level
sync video standards. See
page 192.
the selected Source and Standard:
•
Source channel vertical scaling is set to 140 mV/div.
•
Source channel offset is set to 245 mV.
•
Source channel is turned on.
Trigger type is set to Video.
•
Video trigger mode is set to All Lines (but left unchanged if Standard is
•
Generic).
Display Grid type is set to IRE (when Standard is NTSC) or mV (see
•
select the grid type"
•
Horizontal time/division is set to 10 µs/div for NTSC/PAL/SECAM
standards or 4 µs/div for EDTV or HDTV standards (unchanged for
Generic).
Type
Interlaced/Progressive
Progressive
Progressive
Progressive
Progressive
Progressive
Progressive
Progressive
Progressive
Progressive
Interlaced
Interlaced
"To set up Generic video triggers"
on page 153).
Agilent InfiniiVision 4000 X-Series Oscilloscopes User's Guide
Sync Pulse
Bi-level/Tri-level
Bi-level
Bi-level
Tri-level
Tri-level
Tri-level
Tri-level
Tri-level
Tri-level
Tri-level
Tri-level
Tri-level
on
"To