1 Set up trigger and decode on the serial data signals to be analyzed.
2 Press [Serial] > Lister.
3 Press Window; then, turn the Entry knob to select the size of the Lister
4 Press Display; then, turn the Entry knob to select the serial slot (Serial 1
5 Press the [Single] key (in the Run Control group on the front panel) to
Agilent InfiniiVision 4000 X-Series Oscilloscopes User's Guide
window (Half-Screen or Full-Screen).
When the touchscreen is enabled, you can touch the Lister down or up
chevrons at the top right of the graticule to select the size of the Lister
window.
or Serial 2) on which the serial bus signals are being decoded. (If you
select All, the decode information for different buses is interleaved in
time.)
Before you can select a row or navigate through the Lister data,
oscilloscope acquisitions must be stopped.
stop the acquisition.
7
Serial Decode
145