Trigger Mode/Coupling

To select the Auto or Normal trigger mode

When the oscilloscope is running, the trigger mode tells the oscilloscope
what to do when triggers are not occurring.
In the Auto trigger mode (the default setting), if the specified trigger
conditions are not found, triggers are forced and acquisitions are made so
that signal activity is displayed on the oscilloscope.
In the Normal trigger mode, triggers and acquisitions only occur when the
specified trigger conditions are found.
To select the trigger mode:
1 Press the [Mode/Coupling] key.
2 In the Trigger Mode and Coupling Menu, press the Mode softkey; then
You can also configure the [Quick Action] key to toggle between the Auto
and Normal trigger modes. See
page 347.
Triggering and
After the oscilloscope starts running (after pressing [Run] or [Single] or
the Pre- and
changing the trigger condition), the oscilloscope first fills the pre- trigger
Post-Trigger
buffer. Then, after the pre- trigger buffer is filled, the oscilloscope starts
Buffers
searching for a trigger, and sampled data continues to flow data through
the pre- trigger buffer in a first- in first- out (FIFO) manner.
When a trigger is found, the pre- trigger buffer contains the events that
occurred just before the trigger. Then, the oscilloscope fills the post- trigger
buffer and displays the acquisition memory. If the acquisition was
initiated by [Run/Stop], the process repeats. If the acquisition was initiated
by pressing [Single], the acquisition stops (and you can Pan and Zoom the
waveform).
206
select either Auto or Normal.
You can also make this selection using the touchscreen. See
Trigger Menu, Change the Trigger Mode, and Open the Trigger Level
Dialog"
on page 55.
See the following
"When to Use Auto Trigger Mode"
"When to Use Normal Trigger Mode"
on page 207 descriptions.
"Configuring the [Quick Action] Key"
Agilent InfiniiVision 4000 X-Series Oscilloscopes User's Guide
"Access the
on page 207 and
on