Setting Analog Channel Probe Options

1 Press the probe's associated channel key.
2 In the Channel Menu, press the Probe softkey to display the Channel
See Also

To specify the channel units

1 Press the probe's associated channel key.
2 In the Channel Menu, press Probe.
3 In the Channel Probe Menu, press Units; then, select:
Channel sensitivity, trigger level, measurement results, and math functions
will reflect the measurement units you have selected.
Agilent InfiniiVision 4000 X-Series Oscilloscopes User's Guide
Probe Menu.
This menu lets you select additional probe parameters such as
attenuation factor and units of measurement for the connected probe.
The Channel Probe Menu changes depending on the type of probe
connected.
For passive probes (such as the N2862A/B, N2863A/B, N2889A, N2890A,
10073C, 10074C, or 1165A probes), the Probe Check softkey appears; it
guides you through the process of compensating probes.
For some active probes (such as InfiniiMax probes), the oscilloscope can
accurately calibrate its analog channels for the probe. When you
connect a probe that can be calibrated, the Calibrate Probe softkey
appears (and the probe attenuation softkey may change). See
calibrate a probe"
on page 87.
"To specify the channel units"
"To specify the probe attenuation"
"To specify the probe skew"
Volts — for a voltage probe.
Amps — for a current probe.
on page 85
on page 86
on page 86
3
Vertical Controls
"To
85