To Trigger On Odd Or Even Fields - Agilent Technologies 4000 X-Series User Manual

Infiniivision 4000 x-series
Hide thumbs Also See for 4000 X-Series:
Table of Contents

Advertisement

Figure 31

To trigger on odd or even fields

To check the envelope of your video signals, or to measure worst case
distortion, trigger on the odd or even fields. When Field 1 is selected, the
oscilloscope triggers on color fields 1 or 3. When Field 2 is selected, the
oscilloscope triggers on color fields 2 or 4.
1 Press the [Trigger] key.
2 In the Trigger Menu, press the Trigger softkey; then, turn the Entry knob
3 Press the Settings softkey, then press the Standard softkey to select the
4 Press the Mode softkey and select Field1 or Field2.
Agilent InfiniiVision 4000 X-Series Oscilloscopes User's Guide
Triggering on All Fields
to select Video.
appropriate TV standard.