14
Measurements

Top

The Top of a waveform is the mode (most common value) of the upper
part of the waveform, or if the mode is not well defined, the top is the
same as Maximum. The Y cursor shows the value being measured.
See Also
"To isolate a pulse for Top measurement"
on page 249
To isolate a pulse for Top measurement
The following figure shows how to use Zoom mode to isolate a pulse for a
Top measurement.
You may need to change the measurement window setting so that the
measurement is made in the lower, Zoom window. See
"Measurement
Window"
on page 267.
Figure 46
Isolating area for Top measurement
Agilent InfiniiVision 4000 X-Series Oscilloscopes User's Guide
249