Figure 40
See Also

High Resolution Acquisition Mode

In High Resolution mode, at slower time/div settings extra samples are
averaged in order to reduce random noise, produce a smoother trace on
the screen, and effectively increase vertical resolution.
High Resolution mode averages sequential sample points within the same
acquisition. An extra bit of vertical resolution is produced for every factor
of 4 averages. The number of extra bits of vertical resolution is dependent
on the time per division setting (sweep speed) of the oscilloscope.
The slower the time/div setting, the greater the number of samples that
are averaged together for each display point.
Agilent InfiniiVision 4000 X-Series Oscilloscopes User's Guide
128 Averages used to reduce random noise

Chapter

11, "Trigger Mode/Coupling," starting on page 205
"Averaged Value"
on page 112
12
Acquisition Control
225