14
Measurements
Post Acquisition Processing
N O T E
In addition to changing display parameters after the acquisition, you can perform all of the
measurements and math functions after the acquisition. Measurements and math
functions will be recalculated as you pan and zoom and turn channels on and off. As you
zoom in and out on a signal using the horizontal scale knob and vertical volts/division knob,
you affect the resolution of the display. Because measurements and math functions are
performed on displayed data, you affect the resolution of functions and measurements.

To make automatic measurements

1 Press the [Meas] key to display the Measurement Menu.
2 Press the Source softkey to select the channel, running math function,
3 Press the Type: softkey; then, rotate the Entry knob to select a
242
or reference waveform to be measured.
Only channels, math functions, or reference waveforms that are
displayed are available for measurements.
If a portion of the waveform required for a measurement is not
displayed or does not display enough resolution to make the
measurement, the result will display "No Edges", "Clipped", "Low Signal",
"< value", or "> value", or a similar message to indicate that the
measurement may not be reliable.
measurement to be made.
Agilent InfiniiVision 4000 X-Series Oscilloscopes User's Guide