Csv And Ascii Xy File Structure; Minimum And Maximum Values In Csv Files - Agilent Technologies 4000 X-Series User Manual

Infiniivision 4000 x-series
Hide thumbs Also See for 4000 X-Series:
Table of Contents

Advertisement

CSV and ASCII XY file structure

In CSV or ASCII XY format the Length control selects the number of points
per segment. All segments are contained in the CSV file or in each ASCII
XY data file.
For example: If the Length control is set to 1000 points, there will be 1000
points (rows in the spreadsheet) per segment. When saving all segments
there are three header rows, so the data for the first segment starts at
row 4. The second segment's data starts at row 1004. The time column
shows the time since the trigger on the first segment. The top row shows
the selected number of points per segment.
BIN files are a more efficient data transfer format than CSV or ASCII XY.
Use this file format for fastest data transfer.

Minimum and Maximum Values in CSV Files

If you are running a Minimum or Maximum measurement, the minimum
and maximum values shown in the measurement display may not appear
in the CSV file.
Explanation:
When the oscilloscope's sample rate is 4 GSa/s, a sample will be taken
every 250 ps. If the horizontal scale is set to 10 us/div, there will be
100 us of data displayed (because there are ten divisions across the
screen). To find the total number of samples the oscilloscope will take:
The oscilloscope is required to display those 400K samples using 640 pixel
columns. The oscilloscope will decimate the 400K samples to 640 pixel
columns, and this decimation keeps track of the min and max values of all
the points that are represented by any given column. Those min and max
values will be displayed in that screen column.
A similar process is used to reduce the acquired data to produce a record
usable for various analysis needs such as measurements and CSV data.
This analysis record (or measurement record) is much larger than 640
and may in fact contain up to 65536 points. Still, once the # of acquired
points > 65536, some form of decimation is required. The decimator used
to produce a CSV record is configured to provide a best- estimate of all
the samples that each point in the record represents. Therefore, the min
and max values may not appear in the CSV file.
Agilent InfiniiVision 4000 X-Series Oscilloscopes User's Guide
100 us x 4 GSa/s = 400K samples