I2C Triggering - Agilent Technologies 4000 X-Series User Manual

Infiniivision 4000 x-series
Hide thumbs Also See for 4000 X-Series:
Table of Contents

Advertisement

25
I2C/SPI Triggering and Serial Decode
4 Press the Mode softkey; then, select I2C trigger type.
5 Press the Signals softkey to open the I
6 For both the SCL (serial clock) and SDA (serial data) signals:

I2C Triggering

To set up the oscilloscope to capture I2C signals, see
Signals"
After the oscilloscope has been set up to capture I2C signals, you can
trigger on a stop/start condition, a restart, a missing acknowledge, an
EEPROM data read, or on a read/write frame with a specific device
address and data value.
1 Press [Trigger]; then, select the I2C trigger type.
2 Press [Trigger].
416
a Connect an oscilloscope channel to the signal in the device under
test.
b Press the SCL or SDA softkey; then, turn the Entry knob to select the
channel for the signal.
c Press the corresponding Threshold softkey; then, turn the Entry knob
to select the signal threshold voltage level.
The threshold voltage level is used in decoding, and it will become
the trigger level when the trigger type is set to the selected serial
decode slot.
Data must be stable during the whole high clock cycle or it will be
interpreted as a start or stop condition (data transitioning while the
clock is high).
The SCL and SDA labels for the source channels are automatically set.
on page 415.
Agilent InfiniiVision 4000 X-Series Oscilloscopes User's Guide
2
C Signals Menu.
"Setup for I2C